Abstract:Robot-assisted ultrasound scanning promises to advance autonomous and accessible medical imaging. However, ensuring patient safety and compliant human-robot interaction during probe contact poses a significant challenge. Most existing systems either have high mechanical stiffness or trade performance for compliance. This paper presents a novel compliant end-effector designed to mount on robotic arms for safe and accurate robotic ultrasound imaging, using a quasi-direct drive actuator to achieve passive mechanical compliance and precise active force control. To evaluate the end-effector's performance, we developed an ex vivo dynamic motion simulator platform for contact and scanning testing on tissue under simulated movements. The end-effector was evaluated against a UR3e robot arm using conventional force control strategies as a baseline. Single-point contact experiments from 2.5 N to 15 N show that the end-effector reduced force tracking RMS error by 80.1% on average. Trajectory scanning experiments at different speeds showed an average of 68.0% reduction in force tracking error. Statistically significant improvements were observed in four of six quantitative image quality and stability metrics using the end-effector. This work presents a novel approach for designing and evaluating compliant end-effectors, with the goal of improving safety and reliability in robotic ultrasound.