Abstract:Ultrasound imaging is the most widely adopted medical modality globally due to its low cost and portability, yet artificial intelligence (AI) deployment remains constrained by reliance on GPU-accelerated models, creating a structural paradox where the cost of "intelligence" exceeds that of the imaging device itself. Here, we present the systematic adaptation and extensive evaluation of UltraSeg, an ultra-lightweight architecture originally developed for colonoscopic polyp segmentation, now engineered for point-of-care ultrasound (POCUS) across ten public datasets spanning six anatomical sites (breast, thyroid, kidney, carotid, fetal, and small-animal tumor). We systematically validate both variants in ultrasound domains: UltraSeg-130K (0.13M parameters) achieves 89.7 FPS on single-core CPUs and 34.8 FPS on a refurbished mobile device, while UltraSeg-500K (0.5M parameters) delivers 44.6 FPS on CPU and 16.1 FPS on mobile device. UltraSeg-500K matches or exceeds the Dice performance of the 31M-parameter UNet and approaches 105M-parameter TransUNet in average performance, with superior zero-shot cross-dataset generalization on external validation sets (UDIAT, DDTI). By enabling clinical-grade segmentation without GPU dependency, this work brings AI costs in line with ultrasound accessibility, making advanced diagnostics available in resource-limited settings.